CSS 双腔塔式灭菌器提供多种腔体尺寸,满足不同实验室的空间与装载需求。
EcoCalendar 日历节能
全新灭菌器均配备基于日历的自动开机/关机功能——EcoCalendar。该功能通过软件控制的自动蒸汽关闭程序,帮助降低公用工程消耗和暖通空调(HVAC)负荷。可按一天中的时刻和星期几,对公用工程进行自动启停编程。
EcoJacket 夹套节能
全新灭菌器均标配 EcoJacket,可在灭菌周期之间自动降低灭菌器夹套温度,从而减少蒸汽与水的用量,并降低对房间的热负荷。
Auto Idle Shut-Off 闲置自动关停
与 EcoCalendar 一样,该功能标配于所有全新灭菌器,在设备闲置时切断蒸汽供应以节约能源。其作用类似家用电脑的「睡眠模式」:只需设置期望的闲置时间,灭菌器在闲置时间到后即自动进入睡眠模式。这是节约能源(蒸汽与空调送风)和用水的简便方式,尤其适合间歇使用的灭菌器。
WaterEco 节水系统
所有 Consolidated 灭菌器均配备 WaterEco 系统。除节水外,它还监测用水量,并利用空气、已冷却的排放水与少量冷水的组合为排放水降温。另有兼容冷冻水版本的 WaterEco 系统可供选择。
Hood Control 排风罩联动
该功能将灭菌器与局部排风罩联动,减少对空调送风的无效排放。Form C 干接点信号可在灭菌器门关闭时通知楼宇自控系统关闭排风罩,使实验室更好地控制室内热负荷与环境条件。
Jacket Steam Return 夹套蒸汽回流
使用集中供汽(house steam)的灭菌器可配置夹套蒸汽回流。通过冷凝回水管将夹套中的洁净蒸汽回流至建筑锅炉,帮助节约能源与蒸汽消耗。
| 型号 | 腔体尺寸(宽×高×深) | 容积 |
|---|---|---|
| 2-SSR-2A | 16" x 16" x 26" 40.6 x 40.6 x 66 cm | 3.9 cu. ft 109 liters |
| 2-SSR-3A | 20" x 20" x 38" 50.8 x 50.8 x 96.5 cm | 8.8 cu. ft 249 liters |
Controller Mounting
Every autoclave comes with a control system, however not every control system is mounted in a way that makes the most sense for your laboratory space configuration. Consolidated can mount your control system wherever it best fits your space: right side, left side, top-mounted, or remote-mounted on an adjacent wall.
Components & Plumbing
Consolidated can customize the location of the autoclave’s components and piping to be located on the left or right side of the autoclave, making it easily accessible for you and your service team. This is ideal if you have the need to install your autoclave up against a wall and don’t have access to the autoclave from all sides.
Steam Generator Mounting
If your laboratory is without house steam, Consolidated can provide either an integral (i.e. sits below the chamber) steam generator or a remote-mounted one in a location of your choice (up to 50 feet from the autoclave).
Autoclave Frame Width
If your laboratory has door frames that are smaller than standard, or if your existing wall cutout is narrow, Consolidated can configure an autoclave with a frame dimension that fits through your doorway or existing wall cutout.
Drainage
If your drain is located higher than the flooring, Consolidated can install a sump pump to help transport the sterilizer effluent “uphill” to meet your drain.
Chamber Mounting
If you would like two sterilizers in your laboratory but lack the space required, Consolidated offers a dual “tower” configuration that reduces the footprint by 35% when compared to two autoclaves mounted side-by-side.
Logo Placement
Using our laser cutting equipment, Consolidated can personalize your autoclave by adding your school or company logo to the sterilizer paneling located on the front of the autoclave. This is a big hit for some of our clients!
Custom Loading Carts
If you have odd-sized flasks, beakers, bioreactors or fermenters, Consolidated can customize the size of your loading cart so that it accommodates your load while still fitting into your autoclave chamber.
Additional Shelving
If you’re in need of a more optimized use of your autoclave chamber space, Consolidated can create additional shelving to allow you to efficiently load your autoclave.
Extra Sampling Ports
Consolidated can add additional ports (typically used for validation) with either threaded or sanitary fittings, anywhere on the autoclave chamber as desired.